A Novel Flexible Distributed Pressure Sensing Strip for a Urethral Catheter

Loading...
Thumbnail Image

Persistent link to this item

Statistics
View Statistics

Journal Title

Journal ISSN

Volume Title

Title

A Novel Flexible Distributed Pressure Sensing Strip for a Urethral Catheter

Published Date

2017-06

Publisher

Type

Thesis or Dissertation

Abstract

This dissertation develops, designs, fabricates and evaluates an instrumented catheter for instantaneous measurement of distributed urethral pressure profiles. The developed catheter has important clinical applications in urodynamic testing for analyzing the causes of urinary incontinence in patients. A flexible sensor strip is fabricated with an array of nine pressure sensors and integrated electronic pads for an associated sensor IC chip. The flexible sensor strip and associated IC chip are assembled on a 7 Fr Foley catheter for the urology application. There are two major challenges in the development of the sensor strip. First, a highly sensitive sensor strip that is flexible enough for urethral insertion into a human body is required and second, the sensor should work reliably in a liquid in-vivo environment in the human body. Capacitive force sensors are designed and micro-fabricated using polyimide/PDMS substrates and copper electrodes. To remove the stray influence of urethral tissues which create fringe capacitance that can lead to significant errors, a reference fringe capacitance measurement sensor is incorporated on the strip and the whole sensor area is actively shielded in a Faraday cage made of gold. By supplying the active sensing voltage simultaneously to the deformable electrode of the capacitive sensor and to the Faraday cage, the stray capacitance during in vivo measurements can be largely eliminated. Due to the transparency of the Faraday cage, the top and bottom portions of a capacitive sensor can be accurately aligned and assembled together. Experimental results presented in the thesis show that stray capacitance is reduced by an order of magnitude by the Faraday cage when the sensor is subjected to a full immersion in water. Since the catheter enables a new type of urological measurement, a process for accurate ex-vivo validation of the catheter is also developed. A sheep bladder and urethra are extracted and used in an ex-vivo set up for validation of the developed instrumented catheter. The bladder-urethra are suspended in a test rig and pressure cuffs placed to apply known static and dynamic pressures around the urethra. An algorithm based on use of reference stray transducers is used to compensate for the remaining portion of stray capacitance. Extensive experimental results verify that the developed compensation method works effectively. Results on pressure variation profiles circumferentially around the urethra and longitudinally along the urethra are presented. Finally, the developed sensing catheter is tested in vivo in live female sheep to construct real-time urethral pressure profiles. The results are presented and discussed.

Description

University of Minnesota Ph.D. dissertation. June 2017. Major: Mechanical Engineering. Advisor: Rajesh Rajamani. 1 computer file (PDF); vii, 75 pages.

Related to

Replaces

License

Collections

Series/Report Number

Funding information

Isbn identifier

Doi identifier

Previously Published Citation

Suggested citation

Ahmadi, Mahdi. (2017). A Novel Flexible Distributed Pressure Sensing Strip for a Urethral Catheter. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/190481.

Content distributed via the University Digital Conservancy may be subject to additional license and use restrictions applied by the depositor. By using these files, users agree to the Terms of Use. Materials in the UDC may contain content that is disturbing and/or harmful. For more information, please see our statement on harmful content in digital repositories.