Item Testing the Hydrolysis-Driven Degradation of Sutures: Impacts on Tensile Strength Over Time(2024-12-19) Madhaparia, MihirThis study investigates the rate of tensile strength loss in absorbable sutures due to hydrolytic degradation under simulated in vivo conditions. Sutures were subjected to tensile and degradation testing in a controlled environment to evaluate the impact of moisture and time on suture performance. Results indicate an initial increase in tensile strength upon hydration, followed by a progressive decline due to hydrolysis. These findings provide critical insights for surgeons regarding suture performance in high-stress environments, such as the fascia layer, and underscore the need for further research in dynamic loading and in vivo scenarios.