Abstract

Magnetic Resonance Imaging (MRI) is one of the most sophisticated technologies to produce detailed images of the human body's organs and structures. There were many items of pioneering work in MRI over the last few decades. However, many challenges still remain. This dissertation includes two studies concerning the challenges: the fully automatic matched / tuned RF head coil design and RF coil design with improved B1 field uniformity. Both have been being considered as critical problems at ultra-high field (7T and beyond) MRI systems. The first study aims to build a real-time, high-speed, electrically controlled, and fully stand-alone system of the impedance matched / frequency tuned RF coil that is applicable and compatible to an existing MRI system. To achieve this project, subjects including the basic background of Nuclear Magnetic Resonance (NMR) physics, MRI systems, RF/analog circuit theory. have been studied and presented in this dissertation. The 8-channel RF head coil was successfully built and tested. The fully automatic tuning and matching RF coil offers fast operation (less than 550ms per each channel), accurate impedance matching / frequency tuning (less than -20dB in the reflected coefficient, S11, at the Larmor frequency) resulting in the high power efficiency (4%~21% improved at each channel), and higher Signal-to-Noise Ratio (SNR) in MR images (about 3% in the whole object region). In the second study, the double trapezoidal-like shape along the length of the microstrip resonator is proposed to obtain the gradual impedance variation and flatten the near field distributions. The conventional and proposed 8-channel volume coils have been built and investigated with a phantom in 7T MRI scanner. The results show successful flat field distributions with about 35% increased local transmission magnetic field strength at the ends of the RF coil as well as about 13% improvement at its center. The quality factor ratio between unloaded and loaded is also increased about 45% (from 1.46 to 2.13) compared to the conventional structure. The proposed and demonstrated approaches are a meaningful step in order to overcome difficulties at ultra-high field MRI systems, and one critical contribution to the area.

Description

University of Minnesota Ph.D. dissertation. February 2013. Major: Electrical Engineering. Advisor: Anand Gopinath, Ph.D. 1 computer file (PDF); ii, 119 pages.

Related to

Replaces

License

Collections

Series/Report Number

Funding information

Isbn identifier

Doi identifier

Previously Published Citation

Suggested citation

Sohn, Sung-Min. (2013). RF and electronic design perspective on ultra-high field MRI systems. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/147271.

Content distributed via the University Digital Conservancy may be subject to additional license and use restrictions applied by the depositor. By using these files, users agree to the Terms of Use. Materials in the UDC may contain content that is disturbing and/or harmful. For more information, please see our statement on harmful content in digital repositories.